Instrument products > Mountcorder

MT100

Technical Specs & Details

Specifications

Main Unit Specifications

Item Description
Number of analog input channels 10 channels
External Input/Output Trigger input, Logic input 4 ch or Pulse input 4 ch, Alarm output 4 ch
Sampling interval 100 ms to 1 h (Max:100 ms/10ch)
TIME/DIV 1 sec/DIV to 72 h/DIV
Trigger function Type Start Data capture starts when a trigger is generated
Stop Data capture starts when a trigger is generated
Condition Start Off, Level, Alarm, Scheduled time, External, Day
Stop Off, Level, Alarm, Scheduled time, Time, Day, External
Condition Analog H, L, Window In, Window Out, Rate of change
Logic H, L
Pulse H, L, Window In, Window Out, Rate of change
Alarm output No. of channels 4
Output format Relay contact output (NO/NC)
Output conditions Level judgement, Window judgement, Logic pattern judgement, Pulse judgement
Output terminal M4 screw terminal
Rating 250 VAC/2A, 30 VDC/2A
Pulse input range Count mode 50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M C/F.S. (max. 50 k/sampling interval)
Inst. Mode 50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M C/F.S. (max. 50 k/sampling interval)
RPM mode 50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M RPM/F.S. (max. 50 k/sec)
Calculation function Statistics calculation *1 Average, Peak, maximum, Minimum, RMS (2 calculations can be set simultaneously)
Calculation between channels Addition, Subtraction, Multiplication, Division (for analog input ch1 to ch10)
Wet-and dry-bulb temperature conversion Humidity is calculated from the measured temperature of Wet-bulb and Dry-bulb
(Dry bulb : fixed to ch1, Wet bulb : fixed to ch2, Range 0 to 100 %)
Other function Searching, Annotation input, Message, Marker function
Interface to PC Ethernet (10BASE-T / 100BASE-TX), USB (Compatible with full speed)
Ethernet function Web server function, FTP server function, FTP client function, NTP client function
USB function USB drive mode (File transfer/delete of internal memory)
Internal storage device Internal flash memory : 14 Mbyte, USB memory slot (compatible with full speed) *2
Display Size 5.7 inch TFT color LCD
Displayed items Waveforms + Digital values (vertical, horizontal), Digital + Calculation, Bar graph (vertical)
Clock accuracy (23°C environment) ±0.002 % (within approx. 50 sec per month)
Operating environment Temperature : 0 to 50 °C, Humidity : 5 to 85 % R.H.
Withstand voltage Between each input channel and GND: 1 minute at 350 Vp-p
Between input terminals: 1 minute at 350 Vp-p
Power supply 100 to 240 V AC / 50 to 60 Hz, M4 screw terminal (Power cord is not attached)
Power consumption 38 VA (when LCD is ON)
External dimensions (W x H x D)(approx.) 144 mm x 144 mm x 200 mm
Weight (approx.) 2.1 kg
Vibration resistance  Compatible with JIS Vibration testing methods for automobile Type 1 Class A-equivalent
Dust proof, Drip proof: IP65 compatible (for front panel only)

Analog input Specifications

Item Description
Type of input terminal M4 screw type terminal
Measurement range Voltage 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50, V, 1-5 V F.S.
Temperature Thermocouple : K, J, E, T, R, S, B, N, W (WRe5-26)
R.T.D. : Pt100, JPt100, Pt1000 (IEC751)
Humidity 0 to 100 % RH (Voltage 0 to 1 V conversion, when using optional B-530 humidity sensor is used)
Input filter Off, 2, 5, 10, 20, 40 (Moving average)
Measurement accuracy *3 Voltage ±0.1 % of F.S.
Thermocouple Measurement range Accuracy
R/S 0 TS 100 °C
100 < TS 300 °C
R : 300 < TS 1600 °C
S : 300 < TS 1760 °C
± 5.2 °C
± 3.0 °C
± (0.05 % of rdg + 2.0 °C)
± (0.05 % of rdg + 2.0 °C)
B 400 TS 600 °C
600 < TS 1820 °C
± 3.5 °C
± (0.05 % of rdg + 2.0 °C)
K -200 TS -100 °C
-100 < TS 1370 °C
 
E -200 TS -100 °C
-100 < TS 800 °C
 
T -200 TS -100 °C
-100 < TS 400 °C
 
J -200 TS -100 °C
-100 < TS 100 °C
100 < TS 1100 °C
 
N 0 TS 1300 °C  
W 0 TS 2000 °C  
R.T.D. Pt100 -200 to 850 °C (FS = 1050 °C)  
JPt100 -200 to 500 °C (FS = 700 °C)  
Pt1000 -200 to 500 °C (FS = 700 °C)  
* If the reference junction compensation is internal, add ± 0.5 °C to each of the above values.
A/D converter 16 bit (out of which 14 are internally acknowledged)
Input resistance 1MΩ ±5%
Allowable signal source resistance Up to 300Ω
Maximum input voltage 60 Vp-p (Between ± terminals)
60 Vp-p (Between input terminals)
60 Vp-p (Between input terminal and GND)
Withstand voltage Between each input channel and GND: 1 minute at 350 Vp-p
Between input terminals: 1 minute at 350 Vp-p

Remarks

*1 for realtime and between cursors (when data reviewing)

*2 1 file 2 GB (it depends on used USB memory)

*3 Thermocouple: 0.32φ for T type, 0.65φ for other types

PC software specifications

Item Description
Compatible OS Windows 2000/XP/Vista (32 bit and 64 bit version)
Function Control of main unit, Realtime data capture, File conversion
Max. number of controlable units 10 units
Max. number of controlable channels 100 channels
Settable items AMP settings, Capture settings, Trigger/Alarm settings, Report settings, Other
Captured data Realtime data, Internal memory data, USB memory data (CSV and Binary data)
Display Analog waveforms, Logic waveforms, Pulse waveforms, Digital values
File conversion Data conversion between cursors/All data
Monitor functions Alarm monitor enables sending of email to the specified address
Dual display function Current/previous data display
Statistics calculation, report Maximum, minimum, average data during measurement
Report function Automatic creation of daily or monthly files