BURN OUT” is always displayed even though a thermocouple is connected. 

Remove the thermocouple from the relevant channel and try short-circuiting between the + and – terminals with a short conductor. If the display shows about room temperature (*) after short-circuiting, the thermocouple is disconnected. If “BURN OUT” is still displayed at all times, the main unit may be malfunctioning. 

Please contact Graphtec distributor. 

* The above is the case when pressing [MENU] > “OTHER” > “TEMP. Setting” > ”Room Temp.” is set to “Internal”. 
When “Room Temp.” is set to “External”, approximately 0°C is displayed.