When data is recorded by external sampling, the amplitude of measured voltage and temperature is large or unstable.
If the AC line filter is set to Disable (Off) when external sampling is enabled (On), set it to Enabled (On) and measure at an external sampling interval higher than the fastest interval at which the digital filter is enabled.
The fastest interval is shown in the table below.
Measurement CH* | Fastest interval | |
Digital filter “Off” | Digital filter “On” | |
1CH | 5ms | 200ms |
2CH | 10ms | 500ms |
3 to 4CH | 20ms | 1s |
5CH | 50ms | 1s |
6 to 10CH | 50ms | 1s |
11 to 20CH | 100ms | 2s |
21 to 50CH | 200ms | 5s |
51 to 100CH | 500ms | 10s |
101~200 CH | 1s | 20s |
* The number of measurement channels is the number of channels whose input settings for the standard screw terminal, withstand
high-voltage high-precision terminal, or screwless terminal are set to anything other than Off.